Title Development of a Precise Thermometer for the Measurement of Thermo-physiological Response
Authors 송국섭(Cook-Sup Song) ; 장문녀(Moon-Nyeo Chang) ; 김기훈(Ki-Hoon Kim)
Coverage
(Cover Date)
v.15 n.4(2008-12)
Keywords Thermal comfort ; Skin temperature and skin thermometer
Abstract A skin thermometer is widely used to elucidate the relationship between the physiological response and the surrounding environment. The purpose of this work was to develop a precise skin thermometer with accuracy of ±0.1oC considering to ISO 9886 and ISO 7933. Characteristics of sensors such as RTD pt-100 Ohm, NTC thermistor, Thermocouple and Integrated Silicon Sensor were compared for sensitivity, stability and the availability of an operating amplifier. Finally a 10k-Ohm NTC thermistor was adapted. A ±0.1% resistor was used to evaluate the accuracy of the operating amplifier, and a copper annular with a 50mm-diameter was used to evaluate the characteristics of the thermometer. The accuracy of the thermometer was to be ±0.08oC. The skin thermometer developed in this work was utilized to measure skin temperature for thermo-physiological responses.