Title Gradient Waveform Synthesizer in Magnetic Resonance Imaging System using Digital Signal Processors
Authors 고광혁(Go, Gwang-Hyeok) ; 권의석(Gwon, Ui-Seok) ; 김치영(Kim, Chi-Yeong) ; 김휴정(Kim, Hyu-Jeong) ; 김상묵(Kim, Sang-Muk) ; 안창범(An, Chang-Beom)
Page pp.48-53
ISSN 1975-8359
Keywords Spiral Scan Imaging(SSI) ; Digital Signal Processor(DSP) ; Magnetic Resonance Imaging(MRI) ; Ultra Fast MRI ; Waveform Synthesizer
Abstract In this paper, we develop a TMS320C31 (60MHz) digital signal processor (DSP) board to synthesize gradient waveforms for Spiral Scan Imaging (SSI), which is one of the ultra fast magnetic resonance imaging (MRI) methods widely used. In SSI, accurate gradient waveforms are very essential to high quality magnetic resonance images. For this purpose, sampling rate for synthesizing the gradient waveforms is set twice as high as the data sampling rate. With the developed DSP boards accurate gradient waveforms are obtained. Ultra fast spiral scan imaging with the developed with the developed DSP board is currently under development.