Title Optimum Electrode Selection for Measuring Abdominal Pressure using Bio-Impedance Method
Authors 김거식(Kim, Keo-Sik) ; 이상옥(Lee, Sang-Ok) ; 서정환(Seo, Jeong-Hwan) ; 김경섭(Kim, Kyeong-Seop) ; 송철규(Song, Chul-Gyu)
Page pp.1339-1342
ISSN 1975-8359
Keywords bio-impedance change ; abdominal pressure ; optimum electrode configuration ; urodynamics monitoring
Abstract In this study, we determined the optimum electrode pair for measuring the abdominal pressure using bio-impedance method and compared with conventional methods. Because impedance changes differ from a weight, a height, contractile force, volume of muscle and blood other or whatever of individuals, it was quantified using values of impedance change, correlation coefficient and SNR. Our results showed the optimum electrode pair (1, 9) which could detect impedance changes due to an increase of the intensity of the abdominal pressure. The correlation coefficient and quadratic function between the RMS values of EMG and the impedance changes were 0.87 and y=0.0014x^2+0.0620x+0.6958, respectively. It demonstrated that the abdominal pressure could be measured noninvasively and simply using bio-impedance method. We propose that this optimum electrode configuration would be useful for future studies involving the convenient measurement of abdominal pressure by ambulatory urodynamics monitoring study.