Title Development of 3-Dimensional Polyimide-based Neural Probe with Improved Mechanical Stiffness and Double-side Recording Sites
Authors 김태현(Kim, Tae-Hyun) ; 이기근(Lee, Kee-Keun)
Page pp.1998-2003
ISSN 1975-8359
Keywords Polyimide ; Neural probe ; Double-side recording ; Ni electroplating
Abstract A flexible but implantable polyimide-based neural implant was fabricated for reliable and stable long-term monitoring of neural activities from brain. The developed neural implant provides 3-dimensional (3D) 3×3 structure, avoids any hand handling, and makes the insertion more efficient and reliable. Any film curvature caused by residual stress was not observed in the electrode. The 3D flexible polyimide electrode penetrated a dense gel whose stiffness is close to live brain tissue, because a ~1μm thick nickel was electroplated along the edge of the shank in order to improve the stiffness. The recording sites were positioned at both side of the shank to increase the probability of recording neural signals from a target volume of tissue. Impedance remained stable over 72 hours because of extremely low moisture uptake in the polyimide dielectric layers. At electrical recording test in vitro, the fabricated electrode showed excellent recording performance, suggesting that this electrode has the potential for great recording from neuron firing and long-term implant performance.