Title Optimization of Measuring Cardiac Output by Both Hands Electrode Impedance Method
Authors 정상오(Jung, Sang-O) ; 심명헌(Sim, Myeong-Heon) ; 정운모(Jung, Woon-Mo) ; 김민용(Kim, Min-Yong) ; 윤찬솔(Yoon, Chan-Sol) ; 윤형로(Yoon, Hyung-Ro)
DOI https://doi.org/10.5370/KIEE.2011.60.9.1770
Page pp.1770-1776
ISSN 1975-8359
Keywords ICG(Impedance Cardiography) ; Cardiac output ; Stroke volume ; Non-invasive
Abstract In this study, a new method that can estimate ICG data from a subject's both hands to measure Cardiac Output under convenient sensor environment. With this aim, a grip-type electrode was implemented to measure ICG. To improve the accuracy of measurement, the regression equation was extracted using multiple bio-parameters and our result was compared with the thoracic ICG equipment(Physio Flow^{(R)}, PF104D, Manatec Biomedical, France), which is being used in clinics. The subjects consist of 26 men and 4 women(age in 22.0pm3.32). They are no cardiac disease. Parameters available for regression model were used gender, BMI, MBP, LVET, dZ/dt(max), distance between the measured electrodes, body impedance, and PTT. As a result of analyzing the ICG measurement value on thorax and both hands, the correlation with stroke volume, heart rate, and cardiac output was R^2=0.853, R^2=0.958 and R^2=0.899, respectively.