Title Noise Reduction Method for Particle Measurement System using Beta-ray Absorption Method
Authors 최훈(Choi, Hun) ; 손상욱(Sohn, Sang-Wook) ; 배현덕(Bae, Hyeon-Deok)
DOI https://doi.org/10.5370/KIEE.2012.61.11.1706
Page pp.1706-1712
ISSN 1975-8359
Keywords Noise reduction ; Atmospheric particles ; Beta-ray absorption method ; Linear curve fitting
Abstract The Beta-ray absorption method (BAM) gives a good solution for measuring the mass concentration of atmospheric particles(PM10 and PM2.5). To determine particular matters (PM) concentration, a ratio of the number of detected beta-ray intensity passing through the clean filter and the dust-sampled filter is used. These intensity data measured in air pollution monitoring such as PM10 and PM2.5 usually contained the additive noise(thermal noise, power supply noise and etc.). Therefore, the estimation performance of mass concentration can be deteriorated by these noises. In this paper, we present a new noise reduction method that is essentially required to develope an automatic continuous PM monitoring system using beta-ray absorption method. By combining the block data averaging technique and curve fitting, in the proposed method, the additive noise can be reduced in the measured data. To evaluate the performance of the proposed method, computer simulations were performed with computer generated signals as the input.