Title Measuring Blood Pressure Using Oscillation Signal from an Automatic Sphygmomanometer
Authors 김동준(Kim, Dong-Jun) ; 김영수(Kim, Young-Soo)
DOI https://doi.org/10.5370/KIEE.2012.61.11.1720
Page pp.1720-1724
ISSN 1975-8359
Keywords Blood pressure measuring algorithm ; Automatic sphygmomanometer ; Oscillometric method ; Characteristic ratios method
Abstract This study describes an oscillometric-based blood pressure measuring algorithm by detecting turning points of oscillation signal from digitally filtered cuff signals of an automatic sphygmomanometer. The blood pressure measuring algorithm uses a characteristic ratios method from the turning points. The accurate values of the systolic/diastolic blood presures(SBP/DBP) are calculated using the peaks in the ranges of characteristic ratios. Performances of the proposed algorithm and four automatic sphygmomanometers are compared with the mercury manometer(manual type sphygmomanometer), regarding the SBP and DBP values of manual sphygmomanometer as the reference values. The performance test showed the proposed algorithm revealed the best results in errors and a statistical analysis. Therefore this algorithm can be usable in any automatic sphygmomanometers.ssure states. This may be compromising results for subject-independent sensibility evaluation using EEG signal.