• 대한전기학회
Mobile QR Code QR CODE : The Transactions of the Korean Institute of Electrical Engineers
  • COPE
  • kcse
  • 한국과학기술단체총연합회
  • 한국학술지인용색인
  • Scopus
  • crossref
  • orcid
Title Algorithm to Improve Mass Spectral Resolution of Gas Chromatography Mass Spectrometer
Authors 최훈(Hun Choi)
DOI http://doi.org/10.5370/KIEE.2018.67.9.1232
Page pp.1232-1238
ISSN 1975-8359
Keywords GC-MS ; Mass spectrum ; Curve fitting ; Linear fitting ; Peak position error
Abstract This paper proposes methods for improving mass spectral resolution for a gas chromatograph mass spectrometer. The slope signs of the 1st and 2nd fitting functions for the ion signal block of each mass index are obtained, and the unnecessary element signals in the ion signal block are removed. The spectrum can be obtained by obtaining the second- order fitting function of the reconstructed ion signal block using only the effective ion signals. In addition, the resolution of the mass spectrum can be improved by correcting the error caused by the shift of the spectral peak position. To verify the performance of the proposed methods, computer simulations were performed using the actual ion signals obtained from the GC-MS system under development. Simulation results show that the proposed method is valid.