Mobile QR Code QR CODE : The Transactions P of the Korean Institute of Electrical Engineers
The Transactions P of the Korean Institute of Electrical Engineers

Korean Journal of Air-Conditioning and Refrigeration Engineering

ISO Journal TitleTrans. P of KIEE
  • Indexed by
    Korea Citation Index(KCI)
Title High Frequency Circuit Design using Feedback Control with Body Load Fluctuation for Pain Relief Therapy
Authors 박철원(Park, Chul-Won) ; 원철희(Won, Chul-Hee)
DOI https://doi.org/10.5370/KIEEP.2013.62.1.045
Page pp.45-49
ISSN 1229-800X
Keywords Body load fluctuation ; Deep heat ; Feedback ; High frequency energy ; Medical system ; Push pull switching ; Sinusoidal frequency ; Tailored treatments ; Therapeutic effect
Abstract High frequency system has been used for the purpose of skin care and obesity treatment, by high-frequency energy is applied to the human body generates deep heat. Conventional high frequency system could not selection control by depending on the body load fluctuations. Such as burns and side effects have been reported by system instability and then therapeutic effect is insufficient. During treatment, objective information about the status of the patient was no. Because of treatment methods are subjective, and so tailored treatments were impossible. In this paper, high frequency medical system with sinusoidal frequency characteristics without distortion of the Push pull switching scheme for pain relief therapy was designed. And control circuit that was designed by feedback using the output changes according to the body-load fluctuation. Last, power circuit for efficient control the heat generated from the hardware was proposed.