Mobile QR Code QR CODE : Journal of the Korean Institute of Illuminating and Electrical Installation Engineers

Journal of the Korean Institute of Illuminating and Electrical Installation Engineers

ISO Journal TitleJ Korean Inst. IIIum. Electr. Install. Eng.
Title Development of Spectral Irradiance Measurement System
Authors Jung-Chul Seo
Page pp.48-53
ISSN 1225-1135
Keywords Spectral irradiance ; Spectroradiometry ; Measurement standards
Abstract Spectral irradiance measurement system was developed to measure the spectral irradiance of optical sources in the wavelength range from 250[nm] to 1600[nm]. Our system is composed of source system, fore-optics, monochromator system, optical detector system, and automatic control system. Optical detector system with PMT, Si, InGaAs, and IR enhanced InGaAs detectors is used to measure the wide spectrum of optical sources in ultraviolet, visible, and infrared wavelength regions. Spectral irradiance of the 1[㎾] quartz-halogen tungsten lamp was measured and compared in the wavelength range from 250[nm] to 1600[nm]. The differences between our results and those reported by NIST are below 3[%], 3.5[%], and 5[%] in the wavelength range of 450∼700[nm], 700∼1600[nm], 250∼400[nm], respectively.