Title |
Hybrid Type X-Ray Generator Using EDLC for Fluoroscopy X-Ray System |
Authors |
Young-Min Seo ; Soon-Chan Hong |
DOI |
http://dx.doi.org/10.5207/JIEIE.2014.28.9.089 |
Keywords |
HVPS ; X-ray ; X-ray Generator ; Fluoroscopy ; EDLC ; Energy Storage Device |
Abstract |
A diagnostic fluoroscopy X-ray system uses a 32kW or greater X-ray generator for obtaining real-time moving images and high-resolution images. Fluoroscopy X-ray systems have to use a high-capacity AC power source to perform long-time low-power fluoroscopy and short-time high-power spot exposure. In this paper, we propose a hybrid type X-ray generator for fluoroscopy X-ray system which can perform fluoroscopy and spot exposure with a low-capacity AC power source and an energy storage device. The characteristics of energy storage devices are compared and each energy storage device is modelled to equivalent circuit. And the characteristics of available energy are analyzed as a function of output voltage and power. A 32kW class hybrid X-ray generator with EDLC as an energy storage device for fluoroscopy X-ray system was constructed, and its validity was verified by means of simulations and experiments. |