(Min-Joong Moon)
1
(Kyun-Man Yoo)
2
(In-Seon Yeo)
†
-
(Department of Electrical Engineering, Chonnam National University)
-
(Department of Electrical Engineering, Chonnam National University)
Copyright © The Korean Institute of Illuminating and Electrical Engineers(KIIEE)
Key words
LED Therapy, Measurement Method, LED, Medical Light Source, Irradiance, Beam Angle
1. Introduction
As the global optical medical device market has grown in size, the use of light-emitting
diode (LED) lighting products has steadily increased in both household and hospital
applications (1). However, due to the absence of test standards specifically designed to evaluate
such medical technologies, conventional LED test standards for illumination have been
used instead. A significant challenge with this approach is that conventional standards
were not designed to provide traceability; thus, it is difficult to relate various
research data sets. For this reason, it is essential that studies be conducted on
the test standards used to evaluate medical LED light sources to develop methods to
ensure traceability. Conventional LED test standards, such as CIE 127, KS C 7104,
which specifies a standard method for evaluating the performance of LEDs, and KS C
7652, which specifies the requirements for Non-ballasted LED lamps, primarily focus
on evaluating the performance of the light source itself and detail methods for evaluating
the illuminance of light sources by finding a beam angle at which 50% of the light
intensity is reached at the center point or by measuring the illuminance at a specific
distance or point. Such conventional standards are specifically designed to assess
visible light factors, such as the intensity, luminance, illuminance, and total luminous
flux [2-4]. In contrast, for medical light sources, it is necessary to accurately
analyze the amount of irradiation absorbed by cells and skin tissues instead of measuring
the irradiation required to ensure visibility. Studies on medical light sources require
measurement of the irradiance [mW/㎠] in the development stage in terms of the radiant
energy density [J/㎠] to ensure the optical properties of sufficiently high quality.
As shown in Fig. 1, LED light sources in medical devices are typically tested based on the amount of
light irradiating the cultured cells on the bottom surface of the cell plate. Thus,
an analysis of the optical properties of the light source is required based on the
irradiance that is directly involved in the survival of the cell.
Fig. 1. Mimetic diagram of a cell irradiation experiment
2. Materials and methods
The purpose of the present study is to develop methods to measure the optical properties
of medical LED light sources used in cell testing when their light is directly irradiated
onto a cell plate and to assess the associated impact on cell viability.
Tests were conducted to compare and analyzed the irradiance values calculated for
each measurement method to the results obtained from conventional LED measurement
methods.
All optical measurements were performed in a non-ventilated test environment at a
temperature of (25 ± 3)°C and a relative humidity of (30 ± 10)%, as specified in Annex
A of KS C 7652: 2014 (Method for Measuring Optical Properties of LED Lamps) (5).
The optical equipment used for the measurements was calibrated by a certified calibration
laboratory within one year of the test date.
2.1 LED light source
The LED light source used in the samples was of a type commonly used in therapeutic
applications, such as skin irradiation, and had peak wavelengths between 390 and 698nm
and a light irradiation distance of 1.4cm. It was configured to represent a near-light
source. To verify the reliability of the method used to measure the various properties
of LED light sources, such as the array pitch, light emitting shape, and mixed state,
four types of LED modules were fabricated.
1. The first module consisted of five comb-shaped LED modules with a peak wavelength
of 630nm and a pitch of 2.2cm (Fig. 2).
2. The second consisted of an LED module with two 390nm LED packages (Fig. 3).
3. The third comprised four identical 630nm LED packages (Fig. 4).
4. The fourth consisted of an LED module with one 630nm LED package (Fig. 5).
Fig. 2. Comb-shaped 630nm LED module
Fig. 3. Mixed module with a peak wavelength of 390nm
Fig. 4. LED module with a peak wavelength of 630nm
5. The fifth consisted of an LED module with one 698nm LED package (Fig. 6).
2.2 Measurement system and methods
In this study, comparisons were made between the results obtained using the commonly
used conventional LED test standards (2-7) and those obtained using the four measurement methods evaluated in this study.
Fig. 7. Measurement system: portable optical wave meter(A) and goniophotometer(B, C)
The equipment used in the testing employed a portable optical wave meter and goniophotometer,
both of which are depicted in Fig. 7. The cell plate used for calculating the irradiance was circular with a radius of
2.64cm.
2.2.1 Measurement method based on a portable optical wave meter
The first measurement method was performed as per standard KS C 7612: 1987 (Illuminance
Measurement Method) (2,7) and employed a portable optical wave meter, the specifications of which are shown
in Table 1. The comb-shaped 630nm LED module shown in Fig. 2 was used as the sample for this test, during which the distance between the detector
of the portable optical wave meter and the center of the light source was maintained
at 1.4cm. The irradiance was measured within one minute after power was applied to
the sample (Fig. 8).
Table 1. Specifications of the portable optical wave meter
Equipment name (Model)
|
Portable optical wave meter (HD2102.2)
|
Manufacturer (Nation)
|
Delta Ohm (Italy)
|
Wavelength range
|
400 to 1050nm
|
Wavelength accuracy
|
±1nm
|
Radiant power range
|
0.1mW/㎡ to 2,000W/㎡
|
Test items
|
lux, cd, lux/s, cd/s, W/㎡, mW/㎡, J/㎡, $\mu$J/㎠, $\mu$mol(㎡․s), $\mu$mol/㎡, cd/㎡
|
Fig. 8. Irradiance test using Portable Optical Wave Meter
2.2.2 Measurement method based on the radiant intensity at the center of the light
source
The second measurement method was as per CIE 84 1st Edition (The Measurement of Luminous
Flux) and LM79:2008 (Electrical and Photometric Measurements of Solid-State Lighting
Products) (3-4). The equipment used in this test was a goniophotometer, the specifications of which
are shown in Table 2. The measurements were performed at a distance of 80cm so that the distance between
the detector and the center point of the light source was at least 10 times the size
of the light source. The light source sample was mounted at the center of the jig
and the position of the detector was fixed at 0° on the γ plane and 0° on the C plane.
The radiant intensity (Ic) was measured with respect to the center point of the light
source. The irradiance (Ec) was calculated as follows for a light irradiation distance
of 1.4cm :
Table 2. Specification of goniophotometer
Equipment name (Model)
|
Goniophotometer (Neolight G500)
|
Manufacturer (Nation)
|
PIMACS (KOREA)
|
Wavelength range
|
220 to 1,020nm
|
Wavelength accuracy
|
±0.3nm
|
Radiant power range
|
2 $\times$ 10-8 to 200W/㎡
|
Test items
|
lux, cd, lm, W/㎡, cd/㎡, W, W/sr, nm, beam angle, spatial distribution, spectrum, light
distribution graph, wavelength purity, Vf, If, W, FWHM, XYZ, CIE xy, CCT, CRI
|
2.2.3 Measurement method using the average radiant intensity of the half angle
The third measurement method and associated equipment was the same as that described
in Section 2.2.2 (3-4). In this case, the radiant intensity (Iθ) and light distribution was measured through
a fixed detector when the light source was mounted on a rotating jig in the goniophotometer.
The sample rotated from 0° to 360° in the γ plane in 2° steps and from 0° to 90° in
the C plane in 90° steps. After the measurement was completed, an equation for computing
the half angle (θ) for the light source was derived by considering the radius and
light irradiation distance of the cell plate, as shown in Fig. 9. The half angle is the angle at which light was irradiated into the cell plate in
the cell irradiation test. This was computed to be approximately 62.1° using Eq. (2). However, as the sample in the goniophotometer rotated on the γ plane in 2° steps
and the additional precision is therefore not pertinent, the value was rounded to
62°. After the half angle was identified, the mean value of the radiant intensity
(Iθ) irradiated within 62° of the half angle was found to coincide with the irradiance
computed as per Eq. (3).
Fig. 9. Cell irradiation test
2.2.4 Measurement method using the radiant flux in the half angle
The fourth measurement method employed the method and equipment described in Section
2.2.3. However, the irradiance was calculated as per Eq. (4) using the radiant flux (Wθ) rather than the radiant intensity (Iθ) irradiated within
the half angle.
3. Results and Discussion
The measurement results obtained from each of the previously described measurement
methods with a comb-shaped 630nm LED module having a lambertian light distribution
graph (Fig. 10) as a sample are shown in Table 3. The results for the method that employed the radiant flux in the half angle was
used as the baseline. The variation rate represents the variation rate calculated
via each of the four methods with respect to the baseline value.
When the first measurement method was applied, the average irradiance was found to
be 1.512mW/㎠, which was approximately 44% lower than the baseline value. As shown
in Fig. 11, the comb-shaped 630nm LED module used as a sample emitted light with a pitch of
2.2cm, while the diameter of the light-receiving portion in the detector was relatively
small at approximately 1.6cm. In addition, as the distance between the light source
and detector was fixed at 1.4cm in the tests, a lower amount of radiant flux was received
by the detector, which meant the corresponding irradiance was significantly lower
than that of the other measurement methods.
Fig. 10. Light distribution graph for a comb-shaped 630nm LED module: Comb-shaped
630nm LED module A
(Fig. 2)
Table 3. Irradiance test results of comb-shaped LED module
|
Measurement method
|
Irradiance (㎽/㎠)
|
Average irradiance
(㎽/㎠)
|
Rate of change
(%)
|
1
|
Measurement method based on a portable optical wave meter
|
1.570
|
1.512
|
-44
|
1.480
|
1.450
|
1.550
|
1.510
|
2
|
Measurement method based on the radiant intensity at the center of the light source
|
11.789
|
11.666
|
333
|
11.421
|
11.611
|
11.636
|
11.872
|
3
|
Measurement method using the average radiant intensity of the half angle
|
9.796
|
9.684
|
259
|
9.490
|
9.643
|
9.643
|
9.847
|
4
|
Measurement method using the radiant flux in the half angle
|
2.721
|
2.695
|
Baseline value
|
2.642
|
2.689
|
2.683
|
2.737
|
Fig. 11. Light emission in a comb-shaped 630nm LED module
As shown in Fig. 12, the diameter of the cell plate was approximately 5.28cm while the diameter of the
detector in the portable optical wave meter was approximately 1.6cm. A mismatch of
this magnitude in the relative sizes is problematic and indicates the portable optical
wave meter equipment is actually unsuitable for measuring the amount of radiant flux
actually irradiated onto the cells cultured in the plate in this configuration.
Fig. 12. Size of the cell plate and detector: cell plate(A), and detector(B)
When the second measurement method, which used the radiant intensity (Ic) at the center
of the light source, was evaluated, the average irradiance was found to be 11.666mW/㎠,
which was approximately 7.7 times higher than the irradiance measured using the portable
optical wave meter and approximately 333% higher than the baseline value. The detected
irradiance was higher because the maximum radiant intensity was detected at the center
of the light source due to the nature of the Lambertian LED. In addition, this method
exhibited the highest measurement error. Furthermore, as observed in the measurement
method based on a portable optical wave meter, as this method provided no optical
data that reflected the beam angle irradiated onto the cell plate, it was deemed unsuitable
for measuring the optical properties of a medical light source.
When the third measurement method, which was based on the average radiant intensity
(Iθ) in the half angle, was evaluated and the half angle was applied to the irradiance
calculation, the average irradiance was found to be 9.684mW/㎠. This level was approximately
74% below that of the measurement method based on the radiant intensity at the center
of the light source. However, the irradiance of this third method was approximately
6.4 times higher than that of the measurement method based on a portable optical wave
meter. The difficulty here is that the value of the radiant intensity of radiation
irradiated onto the cell plate versus the angle was computed using the average radiant
intensity to calculate the irradiance, which was obtained from the mixed values between
the radiant intensity irradiated in the downward direction from the center of the
light source and the radiant intensity irradiated according to the angle of incidence.
As a result, the measurement error with respect to the baseline value was found to
be 259%.
The final measurement method, which was based on the radiant flux (Wθ) in the half
angle, was an improvement on the previously described three measurement methods. In
this final method, the horizontal irradiance irradiated onto the actual cells was
computed based on the bottom area of the cell plate. The final average irradiance
was calculated to be 2.695mW/㎠ using the radiant flux (Wθ) within a half angle of
62°, which is the angle at which the light is irradiated onto cells, based on the
21.89㎠ bottom area of the cell plate used in the testing. In this particular test,
a cell plate with a light irradiation distance of 1.4cm and a diameter of approximately
5.28cm was used.
The results of the irradiance test for the LED light sources shown in Figs. 3 to 6
are listed in Table 4. The variation rate in the table indicates the variation rate of the irradiance relative
to the baseline value of each sample. In the measurement method that used the center
of the light source, the variation rate of the irradiance was 343% on average, and
the average variation rate in the measurement method using the average radiant intensity
in the half angle was 263%. When the results in Tables 3 and 4 are compared, the difference
in the variation rate between the measurement methods that used the radiant intensity
of the light source center amounted to approximately 10%, and the difference in the
variation rate between the measurement methods based on the average radiant intensity
in the half angle was approximately 7%. Thus, the error range between the measurement
methods increased as the measurement method approached the conventional test method
used the radiant intensity at the center of the light source.
Table 4. Irradiance test results for various LED modules
|
Measurement method
|
Irradiance (㎽/㎠)
|
Rate of change
(%)
|
Sample
|
1
|
Measurement method based on the radiant intensity at the center of the light source
|
11.571
|
316
|
698nm LED module
|
163.011
|
347
|
Mixed module with peak wavelength 390nm LED
|
29.054
|
345
|
LED module with peak wavelength 630nm
|
26.139
|
374
|
630nm LED module
|
|
Avg.
|
-
|
343
|
-
|
2
|
Measurement method using the average radiant intensity of the half angle
|
9.894
|
256
|
698nm LED module
|
133.147
|
265
|
Mixed module with peak wavelength 390nm LED
|
23.858
|
266
|
LED module with peak wavelength 630nm
|
20.455
|
271
|
630nm LED module
|
|
Avg.
|
-
|
263
|
-
|
3
|
Measurement method using the radiant flux in the half angle
|
2.783
|
Baseline value
|
698nm LED module
|
36.465
|
Baseline value
|
Mixed module with peak wavelength 390nm LED
|
6.522
|
Baseline value
|
LED module with peak wavelength 630nm
|
5.515
|
Baseline value
|
630nm LED module
|
|
Avg.
|
-
|
Baseline value
|
-
|
As shown in Table 5, the sample with the highest content of radiant flux in a half angle of 62° was the
630nm LED module, which also had the narrowest light emitting beam angle and the highest
variation rate of irradiance among the samples. In Table 5, the 698nm LED module has the lowest radiant flux content within the beam angle of
62° and the widest light directivity angle among the light source samples. Therefore,
it was found that the variation in irradiance according to the measuring method was
the lowest light source. This is shown in Table 4.
Table 5. Content rate of radiant flux for LED light sources
NO
|
Sample
|
Content rate of radiant flux
(%)
|
Remarks
|
1
|
698nm LED module
|
77.9
|
Within 62° of the beam angle
|
2
|
comb-shaped 630nm LED module A
|
80.7
|
3
|
comb-shaped 630nm LED module B
|
80.5
|
4
|
comb-shaped 630nm LED module C
|
80.6
|
5
|
comb-shaped 630nm LED module D
|
80.6
|
6
|
comb-shaped 630nm LED module E
|
80.6
|
7
|
Mixed module with peak wavelength 390nm LED
|
82.6
|
8
|
LED module with peak wavelength 630nm
|
82.8
|
9
|
630nm LED module
|
83.4
|
The results of measurements based on the conventional LED measurement method diverged
significantly from the values obtained with the measurement methods evaluated in this
study. The maximum data deviation between the conventional and proposed methods reached
374% because the conventional method was not designed for cell testing applications.
Based on the results, the measurement method based on the radiant flux (Wθ) in the
half angle was found to be the most suitable for testing medical LED light sources
as it was not affected by the beam angle of the light source or the light-emitting
pattern. Furthermore, this method computes the actual irradiated horizontal irradiance
for the cell or strain by considering the situation in which cell irradiation tests
are performed concurrently. In this way, the measurement traceability of this measurement
method can be ensured.
4. Conclusions
The medical LED light source used in this study is of a type commonly used to deliver
irradiation to the human body and for cell testing prior to clinical tests (8-13). For such medical light sources, it is important to measure the irradiance and irradiation
time in order to compute the actual amount of radiation delivered to cells. However,
conventional measurement methods do not measure these properties as their objectives
are to evaluate the performance of the light source itself. Thus, if conventional
measurement methods are used to evaluate medical LED light sources, the resulting
error ranges from 44% to a maximum of 374%. The error range was found to increase
as the beam angle of the light source became narrower or the deviation in the light
distribution became more pronounced. In addition, when the irradiance was calculated
using the radiant intensity rather than the radiant flux irradiated in the cell plate,
the deviation of the irradiance data increased by at least 259%. For this reason,
the conventional measurement method was deemed unsuitable for evaluating medical LED
light sources used in cell testing. Thus, an improved measurement method was required
to calculate the irradiance directly affecting cell viability.
The measurement method based on the radiant flux within the half angle, which was
proposed in this study, calculates the horizontal irradiance involved in cell survival
by using the bottom area of the cell plate as well as the total radiant flux in the
half angle. Considering the rotation condition of the light source in the optical
measurement using the goniophotometer, the total amount of radiant flux irradiated
on the cell plate can be determined by using the half angle. Thus, the traceability
of this measurement method can be ensured regardless of the type of light source.
This measurement method can produce better results in situations where cell culture
test results vary for different medical LED light sources in several studies and prevent
the studies from advancing beyond clinical observation due to the difficulty of validating
certain effects.
Most existing test standards for LED light sources have been standardized for evaluating
the performance of the light source itself (2-7). Thus, when such standards are used in the evaluation of medical LED light sources,
the result is an extremely large error and lack of measurement traceability.
In the future, test standards should be developed for medical LED light sources that
include specialized test standards that consider cell irradiation test conditions.
Acknowledgements
This study was conducted as part of the 2018 Promising Local SME Support Program of
the Industry-Academia Joint Technology Development Project funded by the Ministry
of SMEs and Startups (Project title(Project number): Development of IOT dental massage
device for dental treatment involving multi-wavelength irradiation(S2610990)). We
thank the related government department and Myung Shin Medical Co. Ltd.
References
Barolet D., 2008, Light-Emitting Diodes (LEDs) in Dermatology, Semin. Cutan. Med.
Surg., Vol. 27, pp. 227-238
Goodman T., Heidel G., Muray K., 2007, Measurement of LEDs(CIE 127), Commission Internationale
de l’Eclairage 2nd edn., pp. 16-22
Krochmann J., Reiter H., Vandermeersch F., 1989, The measurement of luminous flux(CIE
84), Comm. Int. de l’Eclairage, 1stedn., pp. 1-50
Dowling K., Ohno Y., Grather M., 2008, Electrical and photometric measurements of
solid-state lighting products(LM-79-08), Illuminating Engineering Society, pp. 1-16
Jang U. J., Kim Y. H., Kim G. I., 2018, Non-ballasted LED lamps(KS C 7652), Korean
Agency for Technology and Standards, pp. 9-16
Yun T. Y., Ko H. S., Ku Y. G., 2018, Standard of measuring the performance of light
emitting diodes(KS C 7104), Korean Agency for Technology and Standards, pp. 1-12
Park S. H., Shin P. S., Lee J. C., 2017, Illuminance measurements for lighting installations(KS
C 7612), Korean Agency for Technology and Standards, pp. 4-8
Wi H. S., 2011, The antifungal effect of light emitting diode on Malassezia yeasts,
Master’s Thesis, Department of Medical Science of Chonnam National University, pp.
5-15
Kim S. M., Lee S. C., Ho Won Y., 2014, Effectiveness of home-use light emitting diode
device for seborrheic dermatitis of the scalp, Korean J. Med. Mycol., Vol. 19, pp.
31-38
Kim S. J., Lee Y. W., 2016, Experimental study on LED irradiating effects on human
cell with the change of the energy and wavelengths, J. Korean Inst. Inform. Technol.,
Vol. 14, pp. 79-84
Kim K. S., Kim S. K., 1987, An experimental study on the effects of low power density
laser on the human gingival fibroblast, J. Oral Med. Pain, Vol. 12, pp. 17-26
Cho Y. J., Dae H. S., 2006, Study of the photoinactivation effect on Propionibacterium
acnes after light irradiation with variable wavelengths, Korean J. Dermatal., Vol.
44, pp. 1332-1338
Hyoung S. K., Chung M. A., Won B. L., et al., 2007, The effect of by LED irradiation
in Archidonic acid-treated human gingival fibroblast, J. Korean Assoc. Oral Maxillofac.
Surg, Vol. 31, pp. 153-158
Biography
He received the M.S. degree in electrical engineering from National Chonnam University,
Gwang-Ju, Korea, in 2014, where he is currently pursuing the Ph.D. degree with the
Department of Electrical Engineering.
he has more than ten years of industrial experience in LED lighting development and
optical measurement.
Since 2011, he has been with the Gwangju Technopark accredited by Korea Laboratory
Accreditation Scheme, Gwangju, Korea, as a Senior Researcher.
His current research interests include optical medical device, sterilization using
UV LED, standardization of optical measurement, and integrated primary optics as well
as biophotonics.
He is currently working on a master's degree in Electrical Engineering at Chonnam
National University, Gwang-Ju, Korea, in 2019.
he has more than ten years of industrial experience in LED lighting development and
optical measurement.
Since 2016, he has been with the Gwangju Technopark accredited by Korea Laboratory
Accreditation Scheme, Gwangju, Korea, as a researcher.
His current research interests include optical medical device using LED, methods of
measuring optical properties.
He received the B.S., M.S. and Ph.D. degrees in electrical engineering from Seoul
National University, Seoul, Korea, in 1979, 1981, and 1989, respectively.
Since 1987 he has been a professor in the Department of Electrical Engineering, Chonnam
National University, Gwangju, Korea.
During 2011 and 2012 he worked as the President of Society of Electro-Physics & Application
in Korean Institute of Electrical Engineers.
His research interests include characterization of SSL, lighting design and implementation
for optimum lighting environment, and graphic rendering of illuminated environments.