Kim Jong Pal 1
-
(School of Mechatronics Engineering, Korea University of Technology and Education,
1600, Chungjeol-ro, Byeongcheon-myeon, Dongnam-gu, Cheonan-si, Chungcheongnam-do,
31253, Korea)
Copyright © The Institute of Electronics and Information Engineers(IEIE)
Index Terms
Electroceuticals, implantable device, neural spike, IA, instrumentation amplifier
I. INTRODUCTION
At the 2018 World Economic Forum, electroceuticals was selected as the next promising
technology that can overcome the shortcomings of chemical drugs (1). The transmission of information from body organs to the brain and control signals
from the brain to body organs are based on digital electrical signals in the form
of spikes through nerves. An electroceuticals is a compound word of electronic and
pharmaceuticals, and is a device that electronically reads and manipulates electrical
spike signals in nerve signals to produce the efficacy of drugs. Fig. 1 shows the configuration of connections between neurons and electroceuticals for manipulation
of nerve signals. In open-loop electroceuticals, it usually consists only of a stimulator
and a digital control unit. Since electronic medicine is inserted and used in the
human body, reliability is very important factor. It is necessary to check whether
the stimulation current is actually injected and the stimulation current threshold
at which nerve signals are activated by the stimulus current. For this reason, a readout
channel (READOUT_CHANNEL) in Fig. 1 is required even in open loop electroceuticals.
Fig. 1. Interface configuration between neuron and electroceuticals.
Fig. 2. Top view of readout channel.
Implantable devices for electroceutical applications require very low power consumption
and a small form factor. The readout channel used to amplify the neural signal must
also have a low power consumption and a small area. Traditional amplifiers have a
closed loop architecture that uses output feedback to the input. The widely used capacitively
coupled IA (CCIA) is a representative example (2-5). The CCIA has precise gain characteristics, but the input and feedback capacitors
for gain and the Miller capacitor for stability occupy area. The open loop architecture
amplifier does not require input/feedback capacitors for gain and does not use feedback,
so no Miller capacitor for stabilization is required. If there is a way to adjust
only the amplification gain, the open loop type amplifier has a fundamental advantage
in terms of area. In the case of an open loop amplifier, the amplification gain distribution
may be larger than that of a closed loop amplifier. However, in measuring the neural
spike signal, the distribution of the amplification gain among chips is not important.
This is because the absolute magnitude of the nerve spike signal does not matter.
Neural information is delivered as the frequency of spikes per unit time. In a closed
loop amplifier, the phase-shifted output signal in the high-frequency region can be
fed back to the input and cause oscillation. For this reason, a large compensation
cap is added to deliberately reduce the bandwidth. Because of this stability compensation,
the bandwidth is lowered, and additional current consumption is required to secure
the desired bandwidth. Therefore, the open loop amplifier fundamentally has a smaller
area and lower current consumption compared to the closed loop amplifier.
In this work, an ultralow power neural amplifier circuit with on-chip spike detection
is presented. For low power and low area, the open loop architecture is selected for
amplifiers, and a method of controlling the input DC level is proposed as a means
for adjusting the amplification gain. In the front-end stage, the gain of the open
loop programmable gain IA (OPGIA) is controlled using the input DC level adjustment.
Also, for the first time, the input DC level is used to adjust the amplification gain
of a multiplier.
Section II shows the architecture of this work. Section III shows the circuit description.
Section IV and V show the measurement results and conclusion.
II. ARCHITECTURE
A top configuration of the neural spike detection circuit is shown in Fig. 2. In order to reduce power consumption, ADC and digital processing method is replaced
by a comparator-based approach for neural spike detection. Two IA (IA1 and IA2) and
nonlinear amplifier (MUL) are placed in series for sufficient gain. DC offset and
low frequency fluctuation are eliminated by high pass filters (HPF1 and HPF2) and
differentiator (DIFF) before IAs and MUL, respectively. The reference voltages (VREF_IA1
and VREF_IA2) of the high pass filters are also supplied to the DC inputs of IAs (IA1
and IA2), respectively, and are used to adjust voltage gain of IAs. The DC output
of the DIFF is generated by the voltage DAC (VDAC3) and is provided to the DC input
of MUL, which can also regulate the amplification gain. The sufficiently amplified
analog spike signal is
Fig. 3. Detail circuit structures of (a) open-loop programmable gain IA (OPGIA), (b)
nonlinear amplifier (NLA), (c) threshold generator (TH_GEN), (d) comparator (CMP),
(e) pulse merger (PM).
converted into a digital signal by a comparator (CMP). The comparison reference signal
of the comparator is generated by the threshold generator (TH_GEN). Even if there
is a single nerve spike, several adjacent pulses may appear at the output of the comparator
due to a nonlinear process, and the output signal of the comparator is digital but
the pulse width is not constant. Pulse merger (PM) combines neighboring pulses that
cannot be physiologically continuously generated for a short time into one pulse,
and makes the pulse width constant. All analog circuits are designed with a supply
voltage of 0.6 V.
III. CIRCUIT DESCRIPTION
In order to use a low operating voltage of 0.6 V, the circuit is designed using a
moderate threshold device instead of a normal threshold device. The threshold voltages
for the moderate threshold PMOS and NMOS are 140 mV and 270 mV, respectively.
Fig. 3 shows the detailed circuit for amplification and filtering. The amplification gains
of both OPGIA and NLA are controlled by the input DC level.
Fig. 3(a) shows the detailed circuit of OPGIA. At IA, the transconductance of the input MOS
(PM1 and PM2) is changed by the input DC value (VREF_IA), and the ac biosignal filtered
by the high pass filter (HPF) is
Fig. 4. Photograph of fabricated READOUT_CHANNEL.
amplified by the transconductance. The high pass filters consist of a capacitor (C$_{HPF}$)
and a switched capacitor-based resistor (R$_{HPF}$). The equivalent resistance (R$_{HPF}$)
for HPF can be freely adjusted with the switching frequency (f$_{sw}$). When a switching
frequency of 100 kHz is applied, the high pass filter has a cutoff frequency of 150
Hz and the equivalent resistance value is 316 M$\Omega$.
IA has a simple structure without no resistor or capacitor array for gain control.
The gain of the IA is controlled by DC level at the input (VM_IA_P and VM_IA_P). The
DC voltage at inputs of IA (VM_IA_P and VM_IA_P) is equal to the reference voltage
(VREF_IA) supplied to the HPF. The DC level for IA gain control is generated by the
voltage DAC(VDAC) with 6 bit resolution. In addition, for IA, a common voltage generator
circuit (VCOM_GEN) is added to prevent amplification failure in some fabrication process
corner conditions. The change in threshold voltage of the NMOS (NM1/NM2) according
to the fabrication process variation is detected and compensated to the common mode
feedback voltage (VCOM). The detailed circuit of VCOM_GEN is depicted in Fig. 3(a).
Fig. 3(b) shows the detailed circuit of NLA. The multiplier gain is also controlled by the
input DC level as in OPGIA. The detailed structures of DIFF and MUL are depicted in
Fig. 3(b) (6). The output DC level of the differentiator is adjusted using voltage DAC (VDAC3)
to obtain the required DC level of the multiplier input. The voltage of VDAC3 is applied
to the reference signal node (VREF_DIFF) in the common mode signal feedback circuit
of the differentiator. The multiplier has a scheme of the gate-bulk input and operates
as a nonlinear amplifier.
The threshold generator (TH_GEN) generates a threshold voltage for the comparator
(CMP) input. TH_GEN takes the output of the NLA as an input and outputs a signal to
be used as a comparison threshold through a level shifter and a low pass filter as
shown in Fig. 3(c).
Fig. 5. Measurement results (a) frequency response example of IA, (b) gain of IA according
to input DC level control, (c) input referred noise, (d) spike detection example with
100 μV input spike, (e)-(h) outputs of MUL according to the MUL input DC levels.
Table 1. Performance summary (measurements)
|
This work
|
TBCAS’17 [7]
|
TBCAS’17 [8]
|
TBCAS’19 [9]
|
Fabrication process [um]
|
0.18
|
0.35
|
0.13
|
0.18
|
Supply voltage [V]
|
0.6
|
3
|
1.2
|
0.5
|
High pass freq. [Hz]
|
195
|
74~750
|
15~232
|
1
|
Low pass freq. [kHz]
|
11
|
6.34~9.4
|
5.2~10.2
|
6.8
|
Bandwidth, BW [kHz]
|
11
|
9.4
|
7.4
|
6.8
|
Amplifier architecture
|
Open loop
|
Closed loop
CCIA
|
Closed loop
CCIA
|
Closed loop
CCIA
|
Gain control means
|
Input DC
|
Cap. ratio
|
Cap. ratio
|
Cap. ratio
|
Input cap./feedback cap. [F]
|
None
|
7.5p/100f
|
30p/125f
|
20p/200f
|
Current consumption, Ic [uA]
|
2.3
|
17.0
|
2.5
|
1.8
|
Readout channel area, A [mm2]
|
0.20
|
0.17
|
0.21
|
0.16
|
Ic × A @BW=10kHz
|
0.43
|
3.07
|
0.72
|
0.41
|
Spike detection domain
|
Analog
|
Digital
|
Digital
|
Analog
|
CCIA: Capacitively Coupled IA
As shown in Fig. 3(d), the comparator used a generally well-known architecture. The comparator output can
often output multiple adjacent pulses for one nerve spike input, which is due to noise
and nonlinear operation through DIFF and MUL. Adjacent multiple pulses that cause
neural spike count error are not simply removed by a hysteresis comparator. Therefore,
an adjacent pulse merging circuit (PM) is designed, which merges the adjacent pulses
within a predefined time window as shown in Fig. 3(e).
IV. MEASUREMENT RESULTS
The neural spike detection circuit is fabricated using the standard 0.18 μm CMOS process.
The main blocks shown in Fig. 2 are marked on the die photo of Fig. 4. The measured transfer function example is shown in Fig. 5(a). The high-pass cut-off and low-pass cut-off frequencies of the transfer function are
195 Hz and 11 kHz, respectively. Fig. 5(b) shows the measured programmable voltage gain of the cascaded OPGIA in accordance
with the voltage DAC (VDAC1) register control. Using the voltage DAC control will
change the input DC level of the IA linearly, and the voltage gain of the IA reaches
a maximum gain of 676 V/V. Fig. 5(c) shows the measured noise spectrum and the integrated input-referred noise is calculated
to be 7.1 μVrms in a frequency range of 100 Hz to 10 kHz.
Fig. 5(d) shows the spike detection results with an input spike amplitude of 100 μV and intervals
of 100 Hz. The positive and negative inputs of the CMP come from the outputs of MUL
and TH_GEN, respectively. As mentioned in the previous section, the gain of the multiplier
can be adjusted by the input DC level. Fig. 5(e)-(g) show the increased multiplier's outputs according to the increased multiplier's inputs.
The correlation between the input DC level and the output peak amplitude of the MUL
is summarized in Fig. 5(h). Supply voltage is 0.6 V, and current consumption of IAs, DIFF, MUL, and CMP is 2
μA, 60 nA, 100 nA, and 210 nA, respectively. The overall power consumption is 1.4
μW. Integrated input-referred noise is measured to be 7.1 μVrms in a frequency range
of 100 Hz to 10 kHz. The measured performance is summarized in Table 1. This work was compared with previous works including CCIA and on-chip spike detection.
All the previous works are equipped with closed loop CCIA type amplifiers, and the
amplification gain is controlled by adjusting the ratio of the input and feedback
capacitors (7-9). To account for area and current consumption at the same time, a single new index
is created by multiplying area and current consumption at the same bandwidth level,
the lower the better. In the case of spike detection in the digital domain, it has
an index value that is improved compared to previous works, and has a comparable level
of the state-of-the-art work in the analog domain.
IV. CONCLUSIONS
This paper presents a neural spike detection circuit, in which all the amplifiers
have an open loop architecture to reduce area and current consumption fundamentally.
A new method for controlling the gain of an open loop amplifier is proposed. The gains
of the OPGIA and multiplier are controlled by a DC input. In addition, the IA incorporates
threshold voltage detection and common mode feedback compensation circuit that can
cope with changes in fabrication process conditions to prevent amplification failure.
The circuit is fabricated using the standard 0.18 μm CMOS process. Changing the DC
level at the IA input has shown that the amplification gain of the IA varies linearly
up to 650 V/V. Amplification gain increase of the multiplier has been observed as
the DC input level increases. It also showed that the spike detection circuit works
well with an amplitude input of 100 µVpk.
ACKNOWLEDGMENTS
This work was supported by the National Research Foundation of Korea (NRF) grant funded
by the Korea government (MSIT) (No. 2020R1F1A1067128, 1711126702).
REFERENCES
, https://www.weforum.org/agenda/2018/09/top-10-emerging-technologies-of-2018/
Deepu C. J., Nov, An ECG-on-chip with 535 nW/channel integrated lossless data compressor
for wireless sensors, Solid-State Circuits, IEEE Journal of, Vol. 49, No. 11, pp.
2435-2448
Wu J., Apr 2016, A 2-µW 45-nV/rtHz readout front end with multiple chopping active-high-pass
ripple reduction loop and pseudo feedback DC servo loop, Circuits and Systems II,
IEEE Transactions on, Vol. 63, No. 4, pp. 351-355
Wang T., Feb 2015, A power-efficient highly linear reconfigurable biopotential sensing
amplifier using gate-balanced pseudoresistors, Circuits and Systems II, IEEE Transactions
on, Vol. 62, No. 2, pp. 199-203
Zhang Jie, June 2018, A low-noise, low-power amplifier with current-reused OTA for
ECG recordings, Biomed. Circuits Syst., IEEE Transactions on, Vol. 12, No. 3, pp.
700-708
Cao W., 2013, Ultra-low-power neural recording microsystem for implantable brain machine
interface, Green Computing and Communications and Internet of Things and Cyber, Physical
and Social Computing, 2013, IEEE International Con¬ference on, 20-23, pp. 1050-1053
Liu Y., Dec 2017, A 64-Channel Versatile Neural Recording SoC With Activity-Dependent
Data Throughput, Biomedical Circuits and Systems, IEEE Transactions on, Vol. 11, No.
6, pp. 1344-1355
Delgado-Restituto M., Apr 2017, System-Level Design of a 64-Channel Low Power Neural
Spike Recording Sensor, Biomedical Circuits and Systems, IEEE Transactions on, Vol.
11, No. 2, pp. 420-433
Kim S., Feb 2019, A Sub-μW/Ch Analog Front-End for Δ-Neural Recording With Spike-Driven
Data Compression, Biomedical Circuits and Systems, IEEE Transactions on, Vol. 13,
No. 1, pp. 1-14
Author
Jong Pal Kim received his B.S. degree in mechanical design from the Department of
Mechanical Design, Chung-Ang University, Seoul, Korea, M.S. degree in mechanical engi-neering
from KAIST, Daejon, Korea, and Ph.D. degrees in electrical engineering and computer
science from Seoul National University, Seoul, Korea, in 1995, 1997, and 2003, respectively.
He was a member of research staff at Samsung Advanced Institute of Technology (SAIT)
from 2001 to 2019.
He was with IMEC at Belgium as a visiting researcher from 2010 to 2012. In 2020,
he joined the Faculty of School of Mechatronics Engineering, Korea University of Technology
and Education, Cheonan, Korea.
His research interests include low power and low noise analog integrated circuits
for biomedical and MEMS applications.